用于创建现代塑料材料的大多数聚合过程都需要含有铝,钛或锌等元素的催化物质。欧洲杯足球竞彩然而,必须最大程度地减少成品中催化剂的存在,以保持其理想的特性。除催化剂外,在生产过程中还使用了许多其他电荷和添加剂 - 增塑剂,润滑剂,稳定剂,中酸盐,中酸盐,抗氧化剂,颜料。这些产品包含元素,例如镁(MG),铁(Fe),铝(Al),磷(P),氯(Cl),钙(CA),铬(CR)或钛(Ti)。
        
          
          
          
          
           
           
           
        It is therefore essential to conduct elemental analyses on the finished polymer to optimize the control over the process. Plastics are generally electrical insulators, acid-resistant and hence difficult to put into solution. This makes X-ray fluorescence the best method to obtain rapid and precise analyses for all the elements of the periodic table, starting from elements of atomic number Z as low as boron (Z = 5).
        Instrument
        对一系列聚丙烯(PP)标准样品进行了测试,以确定Thermo Scientific ARL Perform'X系列光谱仪(图1)用于分析聚合物中的微量元素。
        .png)
        图1。Thermo Scientific ARL PERFORM’X series.
        Results
        在ARL Perform'X系列through optimized coupling between X-ray tube and sample, as well as with the high-power tube and generator. This allows high-sensitivity and lower limits of detection. The comprehensive analytical configuration enables better analysis across the periodic table. Operating the system is extremely simple thanks to the state-of-the art Thermo Scientific OXSAS software and its Analytical Assistant.
        Chemical analyses were offered and used for the calibration of the instrument. Table 1 shows the limits of detection achieved for various elements together with the measurement conditions.
        Table 1.元素和分析条件
        
         
          
           | ELEMENT | 
           4200 W(3 Sigma)[ppm] | 
           2500 W(3 Sigma)[ppm] | 
          
          
           | Mg | 
           0.77 | 
           1。00 | 
          
          
           | al | 
           0.21 | 
           0.27 | 
          
          
           | p | 
           0.12 | 
           0.19 | 
          
          
           | Cl | 
           0.27 | 
           0.35 | 
          
          
           | CA | 
           0.13 | 
           0.16 | 
          
          
           | ti | 
           0.09 | 
           0.12 | 
          
          
           | cr | 
           0.10 | 
           0.13 | 
          
          
           | Fe | 
           0.06 | 
           0.08 | 
          
         
        
        
        在聚合物样品上的痕量元素的精密测试在4200 W上进行。该分析是通过运行样本进行十项连续研究的,计数时间为每元素20秒。结果总结在表2和表3中。
        Table 2.Repeatability for trace analysis in polymers.
        
         
          
           | 跑步 | 
           al[PPM] | 
           CA [PPM] | 
           Ce [PPM] | 
          
          
           | 1 | 
           71.3 | 
           104.4 | 
           24。2 | 
          
          
           | 2 | 
           72。0 | 
           104.2 | 
           24。0 | 
          
          
           | 3 | 
           71.6 | 
           104.1 | 
           23。0 | 
          
          
           | 4 | 
           71.5 | 
           105.1 | 
           23。5 | 
          
          
           | 5 | 
           72。5 | 
           104.9 | 
           23。2 | 
          
          
           | 6 | 
           73。5 | 
           105.2 | 
           23。7 | 
          
          
           | 7 | 
           72。7 | 
           105.6 | 
           23。6 | 
          
          
           | 8 | 
           73。4 | 
           105.8 | 
           22.4 | 
          
          
           | Avg. | 
           72。3 | 
           104.9 | 
           23。5 | 
          
          
           | SD | 
           1。4 | 
           1。0 | 
           1。3 | 
          
         
        
        
        表3。聚合物中痕量分析的重复性(4200 W)。
        
         
          
           | 跑步 | 
           Fe [ppm] | 
           Mg [PPM] | 
           ti[PPM] | 
           p[PPM] | 
          
          
           | 1 | 
           11。4 | 
           60.1 | 
           2。1 | 
           15。6 | 
          
          
           | 2 | 
           11。6 | 
           59.9 | 
           2。0 | 
           15。6 | 
          
          
           | 3 | 
           11。6 | 
           60.8 | 
           2。0 | 
           15。6 | 
          
          
           | 4 | 
           11。4 | 
           60.6 | 
           2。1 | 
           15。7 | 
          
          
           | 5 | 
           11。6 | 
           61。3 | 
           2。0 | 
           15。6 | 
          
          
           | 6 | 
           11。4 | 
           61。5 | 
           2。0 | 
           15。9 | 
          
          
           | 7 | 
           11。6 | 
           61。7 | 
           2。2 | 
           16。1 | 
          
          
           | 8 | 
           11。6 | 
           60.5 | 
           2。1 | 
           16。2 | 
          
          
           | Avg. | 
           11。5 | 
           60.8 | 
           2。0 | 
           15。8 | 
          
          
           | SD | 
           0.1 | 
           0.28 | 
           0.1 | 
           0.4 | 
          
         
        
        
        Conclusion
        Analysis of these elements at low concentration levels require high sensitivity and accuracy. These results show that theARL Perform'X系列适用于在聚合物样品中检测Mg,Al,P,Cl,Fe,Ca和Ti的高精度结果。其他元素本来可以被同样鉴定出来,但在用于该测试的标准样本中不存在。由于最佳的电源管理,ARL Perform'X光谱仪可以在2500 W处运行,而无需外部水冷却。因此,在这些情况下,不需要自来水,也不需要水冷却器。在较高的功率水平(4.2 kW)下,由于X射线管功率的智能管理,获得了X射线管的节能和减轻压力。此外,Windows 7下的最先进的OXSA分析软件提供了详细的功能和易用性。
        
        This information has been sourced, reviewed and adapted from materials provided by Thermo Fisher Scientific - Elemental Analyzers.
        For more information on this source, please visitThermo Fisher Scientific-元素分析仪。